MRI tests safe for people
with implanted cardiac devices when certain guidelines are followed
Magnetic Resonance Imaging (MRI), an important
diagnostic test, has traditionally been off limits to more than
2 million people in the United States who have an implanted pacemaker
to regulate heart rhythms or an implanted defibrillator to prevent
sudden cardiac death. Now, in a study published in the October 4
issue of Annals of Internal Medicine, cardiologists at Johns Hopkins
report that a protocol they developed has proved effective in enabling
patients with implanted cardiac devices to safely undergo an MRI
scan.
"We believe this is the largest prospective study of MRI in
patients with implanted devices," says lead author Saman Nazarian,
a Johns Hopkins cardiac electrophysiologist and an assistant professor
of medicine at the Johns Hopkins University School of Medicine.
"The guidelines we have published can be used to make MRI
more available to people who could benefit from early detection
of cancer and other diseases and for guiding surgeons during procedures.
MRI is considered superior to CT scans in many clinical scenarios,
especially for brain and spinal cord imaging," adds Nazarian.
To date, more than 700 patients with implanted cardiac devices have
safely undergone MRI exams at Johns Hopkins.
Their prospective study followed 438 people with implanted cardiac
devices who had 555 MRI scans. Almost all of the exams, 94 percent,
were conducted at The Johns Hopkins Hospital. The rest were performed
at Rambam Medical Center in Haifa, Israel. The researchers found
that with appropriate precautions, patients with pacemakers and
defibrillators can have an MRI scan with very low risk of the device
malfunctioning, moving, heating or causing abnormal heart rhythms
due to the magnetic and radiofrequency energy generated by the test.
The devices implanted in three of the patients in the study, 1.5
percent, had a power-on reset event during an MRI scan, which means
that the energy emitted from the scanner caused the devices to revert
to default settings. This is a rare occurrence that warrants close
expert monitoring during the test, but is easily remedied after
the test is completed. None of the three had device dysfunction
during the long-term follow-up of between 15 and 66 weeks. One of
those patients completed four repeated MRI examinations during the
study without any problems.
Johns Hopkins cardiac electrophysiologist and biomedical engineer
Henry Halperin began researching the issue of MRI safety with implanted
devices about 15 years ago, testing a range of devices. The safety
protocol he developed is now being adopted by institutions around
the world.
"The newer pacemakers made after 1998 and defibrillators manufactured
since 2000 come with electromagnetic interference protection,"
says Halperin, who is a professor of medicine at the Johns Hopkins
University School of Medicine and the senior author of the study.
In addition to the age of the device, the Hopkins team checks the
type and configuration of the leads attached to the device. For
example, if a lead is disconnected and is not part of the device's
function, an MRI would not be recommended because the tip of the
wire could get very hot.
"We reprogram the device to a safe mode while the patient
is having the MRI scan," says Rozann Hansford, R.N., M.P.H.,
a study author who monitors patients at Johns Hopkins during the
scans. "We carefully monitor the patient's blood pressure,
electrical activity of the heart and oxygen saturation, and look
for any unusual symptoms. After the test, we reprogram the device
and carefully check its function." The patients' devices are
checked again in 3 to 6 months.
The researchers conclude that with a protocol based on device selection,
programming and careful patient monitoring, MRI can be performed
safely in many patients who have a pacemaker or a defibrillator.
"With the advancing age of the population and the expanding
indications for pacemakers and defibrillators, this has become an
increasingly important issue, and a lifesaving one for some patients,"
says Nazarian. He adds that many of the patients with cardiac devices
who have come to Hopkins for an MRI scan had tumors and other serious
problems diagnosed and treated, whereas those problems had been
missed by a previous imaging test, such as a CT or ultrasound exam.
In addition to Nazarian, Halperin and Hansford, other Johns Hopkins
authors on the study were Menekhem M. Zivman, Ph.D., Albert C. Lardo,
Ph.D., Brian S. Caffo, Ph.D., Kevin D. Frick, Ph.D., Michael A.
Kraut, M.D., Ph.D., Ihab R. Kamel, M.D., Ph.D., Hugh Calkins, M.D.,
and Ronald D. Berger, M.D., Ph.D. At Rambam Medical Center in Haifa,
Israel, were Ariel Roguin, M.D., and Dorith Goldsher, M.D., and
at NIH, David A. Bluemke, M.D., Ph.D.
|